Model 14 Oxygen Analyzer
6.1.1. Automatic Temperature Compensation
The sensor continuously measures the sample temperature. The concentration values are automatically
calculated using the algorithms programmed in the transmitter.
6-2
TEMP. CALIB.
TEMP. CALIB.
CALCAL
21.2 21.2
˚C ˚C
6.87ppm
OK
Press OK when the
measurement is stable
Can change the value
EXECUTION
T: 0.0 ˚C
ENTER : ok
ESC : cancel
Select
21.2 ˚C
Enter
Enter
1. Programming
Choose automatic temperature compensation. ?
TEMP. COMP.
MEASURE
Select
Select
P: 0760.0 mm Hg
TEMP. COMP
SALINITY COMP.
TYPE: Auto.
Enter
PROGRAMMING
Select
MEASURE
ALARMS
mA OUTPUTS
RS485
Enter
2. Execution
The execution is begun in the CALIBRATION
menu.
Enter the sample temperature value in °C.
Press OK to adjust the temperature displayed
to the actual value of the sample temperature
measured with a precise thermometer.
The gap between the calibration and the
theoretical response curve of the sensor is
given for information.